Model Number 2090
Device Problems Electromagnetic Interference (1194); Difficult to Interrogate (1331); Telemetry Discrepancy (1629)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported that problems with device interrogation with the programmer and radio-frequency (rf) head were occurring; there were consistent problems with gaining access to telemetry.The device nurse indicated there was an issue with electro-magnetic interference and inability to gain telemetry access at this office location.Of note, multiple devices were successfully interrogated for demonstrative purposes with this programmer and rf head.The rf head was changed out to complete the testing and a new rf head was ordered.The rf head was returned for repair.At this time the programmer remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4829456
MDR Text Key18628768
Report Number2182208-2015-01899
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/28/2012
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/06/2015
Initial Date FDA Received06/09/2015
Date Device Manufactured02/25/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
2067 RADIO FREQUENCY HEAD
Patient Age00089 YR
-
-