Device Problem Migration or Expulsion of Device (1395)
Patient Problem Failure of Implant (1924)
Event Date 06/03/2015
Event Type  Injury  
Event Description
Medtronic (covidien) received report that a patient underwent pipeline flow diversion treatment; it was discovered at the 6-month follow-up that the device had migrated proximally.The aneurysm was unruptured and saccular located in the right ophthalmic artery.As a result of the device migration, a pipeline flex was placed to cover the neck of the aneurysm.Immediate angiographic result was slow flow and eclipse.The second pipeline was implanted successfully.No patient injury was reported as a result of this procedure.
 
Manufacturer Narrative
The lot history record review was not possible since the lot number was not reported.The device will not be returned for analysis as it was discarded; therefore, the event cause could not be determined.Reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE EMBOLIZATION DEVICE
Type of Device
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
COVIDIEN (IRVINE)
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
COVIDIEN (IRVINE)
9775 toledo way
irvine CA 92618
Manufacturer Contact
tricha miles
9775 toledo way
irvine, CA 92618
9496801224
MDR Report Key4868884
MDR Text Key18040828
Report Number2029214-2015-00728
Device Sequence Number1
Product Code OUT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100018.S004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/03/2015
Initial Date FDA Received06/25/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age31 YR
-
-