|
Model Number ORCHESTRA PLUS |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Issue With Displayed Error Message (2967)
|
Patient Problem
No Information (3190)
|
Event Date 05/27/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that during a pacemaker check, successful interrogation and tests were performed.However, a freeze occurred on a message.Upon rebooting the subject programming, the following message ¿head successfully initialised¿ was displayed and it was not possible to interrogate the pacemaker.The programmer on/off button was pressed several times and also the power cable was pulled out several times; however, the same message was displayed and it was still unable to interrogate the pacemaker.An explanation is required.
|
|
Event Description
|
It was reported that during a pacemaker check, successful interrogation and tests were performed.However, a freeze occurred on a message.Upon rebooting the subject programming, the following message ¿head successfully initialised¿ was displayed and it was not possible to interrogate the pacemaker.The programmer on/off button was pressed several times and also the power cable was pulled out several times; however, the same message was displayed and it was still unable to interrogate the pacemaker.An explanation is required.
|
|
Search Alerts/Recalls
|
|
|