Model Number MODEL 250
Device Problem Failure to Interrogate (1332)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/23/2015
Event Type  malfunction  
Event Description
The nurse reported that she was unable to advance to the interrogation screen from the main menu.She was unable to advance past the first screen on her handheld.During the report, the screen lock was switched to the lock position and then back to the unlock position.A hard reset was attempted, and the edges of the screen were cleaned.Another hard reset was performed.It was confirmed that the stylus was being used, but was still unable to advance past the screen.The handheld and related software were received by the manufacturer for analysis.However, analysis has not been completed to date.
 
Event Description
Analysis was completed on the handheld and software, and the reported allegation was confirmed.The cause for the display anomaly is associated with resistance values being higher than expected in the touch screen circuitry.Once the display was replaced with a known good display, no further anomalies associated with the handheld performance were identified during the analysis.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4913626
MDR Text Key6030069
Report Number1644487-2015-05190
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1075845
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 06/23/2015
Initial Date FDA Received07/14/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/28/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-