The nurse reported that she was unable to advance to the interrogation screen from the main menu.She was unable to advance past the first screen on her handheld.During the report, the screen lock was switched to the lock position and then back to the unlock position.A hard reset was attempted, and the edges of the screen were cleaned.Another hard reset was performed.It was confirmed that the stylus was being used, but was still unable to advance past the screen.The handheld and related software were received by the manufacturer for analysis.However, analysis has not been completed to date.
|
Analysis was completed on the handheld and software, and the reported allegation was confirmed.The cause for the display anomaly is associated with resistance values being higher than expected in the touch screen circuitry.Once the display was replaced with a known good display, no further anomalies associated with the handheld performance were identified during the analysis.The flashcard and software performed according to functional specifications.
|