Siemens healthcare diagnostics has determined that dimension urine opiate screen operated according to specification; siemens is filing this only to report a patient death at a subsequent treatment facility.
The urine opiate screen detects morphine.
Although heroin is metabolized to morphine, the concentration of morphine in urine may not be above the cut-off concentration (300 ng/ml) of the assay.
As stated in the intended use of the dimension flex reagent cartridge urine opiates screen instructions for use: "opiates are absorbed rapidly.
Heroin is converted almost immediately to morphine, which is excreted in urine both unchanged and as a glucuronidated metabolite.
Excretion takes place over a period of a couple of days.
" naloxone was administered in the ambulance which indicates that they were following standard procedure for a heroin overdose; ongoing management of the patient would require titration of naloxone as needed to maintain breathing and consciousness.
Additionally, the dimension urine opiates screen states in the instructions for use: "the opi method provides only a preliminary analytical test result.
A more specific alternate chemical method must be used in order to obtain a confirmed analytical result.
Gas chromatography/mass spectrometry (gc/ms) is the preferred confirmatory method.
Other chemical confirmation methods are available.
Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
" diluting a urine sample and running the sample and multiplying by the dilution factor is non-standard use when the urine opiate screen is run in qualitative mode.
Per the instructions for use, the dilution can only be used in the semi-quantitative mode.
|