• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37603
Device Problems Migration or Expulsion of Device (1395); Unstable (1667)
Patient Problem Shaking/Tremors (2515)
Event Date 06/29/2015
Event Type  Injury  
Event Description
It was reported there was lead migration.Surgical intervention was required as a result of the event and it was replaced with a new lead and new ¿stim¿ cap.The issue was resolved.The cause of the issue was determined ¿ the stimloc ¿pac-man¿ slipped and didn¿t lock in place which caused the lead to move.The patient didn¿t have tremor control and they had tremor in their right hand.Their status at the time of report was alive with no injury.
 
Manufacturer Narrative
Concomitant products: product id: 3387s-40, lot# va0urxw, explanted: (b)(6) 2015, product type: lead.Product id: 37642, serial# (b)(4), product type: programmer, patient.Product id: 3708660, serial# (b)(4), implanted: (b)(6) 2015, product type: extension.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4919648
MDR Text Key16162629
Report Number3004209178-2015-13549
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/14/2016
Device Model Number37603
Device Catalogue Number37603
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/29/2015
Date Device Manufactured03/25/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age00084 YR
-
-