Catalog Number 0940035000 |
Device Problem
Dull, Blunt (2407)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/24/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that the blade went dull after removing 3 casts.It was also reported that when trying to remove the fourth cast, it was noted that the teeth on the blade were worn down.It was also reported there were no adverse consequences as a result of this event, the procedure was completed successfully without delay.
|
|
Manufacturer Narrative
|
A follow up report will be filed once the quality investigation is complete.Device not yet returned to manufacturer.
|
|
Manufacturer Narrative
|
A dhr (device history record) review of the manufacturing data during the manufacture of this blade indicated that the blade met all manufacturing specifications.Measurements performed on the returned blade against part print specification indicated that the blade met all measurements performed.
|
|
Event Description
|
It was reported that the blade went dull after removing 3 casts.It was also reported that when trying to remove the fourth cast, it was noted that the teeth on the blade were worn down.It was also reported there were no adverse consequences as a result of this event, the procedure was completed successfully without delay.
|
|