Model Number 8700-0752-01 |
Device Problems
Break (1069); Detachment Of Device Component (1104); Battery Problem (2885)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/01/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product in complaint was returned to zoll on (b)(6) 2015 for investigation.However, investigation is still in progress.A supplemental report will be filed once investigation has been completed.
|
|
Event Description
|
It was reported that the latch of the autopulse li-ion battery broke off.No adverse patient sequelae was reported.No further information was provided.
|
|
Manufacturer Narrative
|
Autopulse li-ion battery (s/n (b)(4)) was returned to zoll (b)(4) for evaluation.A visual inspection was performed and observed the battery latch to be damaged/broken.The customer's reported complaint was confirmed.The root cause of the reported problem of the latch being broken elease could not be determined.However, this could have resulted from excessive force during handling of the battery placement in to the autopulse platform.
|
|