Model Number 2090 |
Device Problems
Crack (1135); Noise, Audible (3273)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/19/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is based solely on device analysis.No information to suggest a device related adverse event or product problem was received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: analysis confirmed the customer comment that the overlay was cracked, power cord bay door broken, one hinge plate screw missing and system fan was noisy.Analysis also found that the electrocardiogram connector was loose and therefore the link electronic module printed circuit board was replaced and calibrated.All found defective parts were replaced.Product id: 229047 software analyzer.(b)(4).
|
|
Event Description
|
It was reported that the programmer screen was cracked, the power cord lid was broken, a screw came out of the bottom of the programmer and the fan was loud.The programmer was returned for repair.No patient complications have been reported as a result of this event.It was further reported that the programmer subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
|
|