It was reported that during an transfemoral case in the left internal carotid artery, the physician had significant trouble retracting the blue slider.The stent deivery system experienced difficulty deploying the stent.The stent deployed successfully, and the procedure was completed without any patient harm.
|
|
Device investigation: the device was not returned to inspiremd for investigation, however, production records review confirmed the device met all manufacturing specifications, with no malfunction or nonconformance identified.An investigation was initiated to identify the underlying failure mode(s) associated with delivery system deployment complications which may result in high resistance or inability to deploy the stent.This mdr is being submitted retrospectively as part of corrections implemented through cap-us-006-26 to ensure reporting compliance.At the time of the event, the complaint was assessed as non-reportable; however, the retrospective review determined that the event meets the reporting requirements of 21 cfr part 803 since the reported event may cause or contribute to an adverse event if it were to recur.Inspiremd's recall (recall no.Z-2330-2026), initiated on may 1, 2026, identified a delivery system issue in the cguard® prime carotid stent system.However, this event occurred prior to the initiation of the recall activities.Capa-027-25 has been initiated to further evaluate similar field-reported performance issues, and complaint trending activities remain ongoing.
|