Allergan (Costa Rica) Inspira Smooth Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled Back to Search Results
Catalog Number SRM-330
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 05/13/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported "capsular contracture (baker grade 3)".This record is for the right side.Device has been explanted.
 
Additional Manufacturer Narrative
The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.A review of the device history record has been completed.No deviations or non-conformances noted.This report was impacted by emdr scheduled maintenance occurring july 22-27, 2026.
 
Event or Problem Description
Healthcare professional reported "capsular contracture (baker grade 3)".This record is for the right side.Device has been explanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: d9, h3, h6.Lab results: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (creases and deformation) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inspira Smooth Silicone Gel Filled Breast Implant
Common Device Name
prosthesis, breast, noninflatable, internal, silicone gel-filled
Manufacturer (Section D)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
+1(847)936-6324
MDR Report Key50008360
Report Number9617229-2026-14002
Device Sequence Number12864835
Product Code FTR
UDI-Device Identifier10888628007093
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSRM-330
Device Lot Number1279334
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/14/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 06/26/2026
Supplement Date Received by Manufacturer09/28/2026
Initial Report FDA Received Date07/27/2026
Supplement Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexFemale
Patient Weight125 LB
-
-