• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Encore Medical L.P. DJO SURGICAL; INSERT, ONLAY TIBIAL, EPIK, SIZE40X9

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Encore Medical L.P. DJO SURGICAL; INSERT, ONLAY TIBIAL, EPIK, SIZE40X9 Back to Search Results
Catalog Number 340-40-009
Medical Device Problem Codes Unstable (1667); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526)
Date of Event 07/06/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to pain and instability.
 
Additional Manufacturer Narrative
The agent reported, "patient presented with instability and pain.Male 58".The previous surgery and the surgery detailed in this event occurred 11 years, 27 days apart.Item number: 340-40-009."item description: insert, onlay tibial, epik, size40x9".Lot#: 328b1011.Part revision: a.Product type: knee.Manufacture date: 23-oct-2014.Expiration date: 28-oct-2019.This evaluation is limited in scope as the item(s) associated with this investigation was not returned to (b)(6) for examination.The surgery was completed as intended.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Root cause: the root cause of this complaint was revision surgery for instability and pain.There are multiple factors that may contribute to this event that are outside of the control of enovis surgical.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review a review of the device history record(s) shows that the reported component(s) used in the previous surgery met design and manufacturing requirements at the time of release for use.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported device(s) showed no present trends or ongoing issues that need review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
INSERT, ONLAY TIBIAL, EPIK, SIZE40X9
Manufacturer (Section D)
Encore Medical L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
Encore Medical L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
+1(512)834-6313
MDR Report Key50010815
Report Number1644408-2026-01418
Device Sequence Number14913151
Product Code HRY
UDI-Device Identifier00888912116022
UDI-Public00888912116022
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K020741
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/28/2019
Device Catalogue Number340-40-009
Device Lot Number328B1011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2026
Initial Report FDA Received Date07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/23/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
200-01-110 Lot: 679A1003.; 300-01-040 Lot: 068B1008.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR