|
(b)(4)
date sent 8/17/2026.D4 batch # a98z0f.Corrected data d4: udi#
investigation summary.The product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the 5dcs device was returned with no apparent damage.In addition, the tyvek was returned along with the instrument.To evaluate the reported cutting issue, the device was tested for functionality and successfully cut the test media as intended, with no anomalies observed during testing.The device was found to be fully functional and operating in accordance with manufacturing specifications.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device was returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch a98z0f number, and no non-conformances were identified.
|