Abbott Medical FlexNav Delivery System; Aortic Valve, Prosthesis, Percutaneously Delivered
|
Back to Search Results |
|
| Model Number |
FNAV-DS-LG |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a navitor valve was selected for implant using a large flexnav delivery system (ds).Pre-implant balloon valvuloplasty (pre-bav) was performed using an unspecified balloon.The same wire was reported to be used for pre-bav, pacing, and the valve implant procedure.During the procedure, the valve was able to be implanted.Upon removal of the ds, it was difficult to remove, and it was removed together with a xs, circulo guidewire.The patient remained hemodynamically stable throughout the procedure and there was no clinically significant delay reported.The patient was reported to be discharged.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of stuck guidewire in the delivery system was reported.The investigation into the returned device confirmed the stuck guidewire in the delivery system.The radiopaque tip and valve capsule were observed to contain bent damage, which is consistent with damage during use.The inner shaft was noted to have an accordion damage, likely a result of the attempts to remove the guidewire.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on the information received, the cause of the reported event could not conclusively be determined.However, an exception was initiated to evaluate the reported issue.The investigation identified the root cause as variability in guidewire coil behavior, which can interfere with flexnav delivery system crimps¿leading to advancement or removal issues.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
|
| |
|
Search Alerts/Recalls
|
|
|