| Model Number |
CDS0706-XTW |
| Medical Device Problem Code |
Incomplete Coaptation (2507)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Mitral Valve Insufficiency/ Regurgitation (4451)
|
| Date of Event |
06/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, a patient presented with grade 4 degenerative mitral regurgitation (mr) for a mitraclip procedure.The final mr was grade 1.There were no adverse patient effects or clinically significant delays reported.\
\
on (b)(6) 2026, patient returned with declined ejection fraction, increase in fluid retention, shortness of breath.Transesophageal echocardiogram (tee) was performed, and it was determined that there was a single leaflet device attachment (slda) and the clip was no longer attached to the posterior leaflet.Mr was moderate to severe after slda.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, a patient presented with grade 4 degenerative mitral regurgitation (mr) for a mitraclip procedure.The final mr was grade 1.There were no adverse patient effects or clinically significant delays reported.\
\
on (b)(6) 2026, patient returned with declined ejection fraction, increase in fluid retention, shortness of breath.Transesophageal echocardiogram (tee) was performed, and it was determined that there was a single leaflet device attachment (slda) and the clip was no longer attached to the posterior leaflet.Mr was moderate to severe after slda.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.\
based on the information reviewed, a cause of the reported single leaflet device attachment (slda) resulting in mitral regurgitation (mr) could not be determined.Dyspnea is a cascading effect of the recurrent mr.\
the reported patient effects of mitral regurgitation (mr) and dyspnea are known possible complications associated with mitraclip procedures.The reported serious/injury/ impairment was a result of case-specific circumstance.There is no indication of a product issue with respect to manufacture, design or labeling.
|
| |
|
Additional Manufacturer Narrative
|
|
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|