| Model Number |
OPTI4510CSF10 |
| Medical Device Problem Code |
Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that:
"the physician planned to treat acom aneurysm and primed the vasco 10 micro catheter after prechecked the optima complex 4.5x10 coil he placed it inside the aneurysm and tried to detach the coil.After multiple attempts and tried all the techniques the coil didn't detach so he tried to resheath the coil into the micro catheter the coil got detached halfway into the fargo guiding catheter.The physician removed the coil by aspirating and used another optima to finish the procedure."
update 29jun2026 - additional information received from the issuer:
"1.The vessel was torturous.2.The detacher, micro catheter is not available for return.3.Precheck was performed on the coil.4.The xcel showed red and green light when the physician was ready to detach.We tried rotating the xcel, wiping the detachment gold marker end, still red and green light was blinking."
incident report form submitted for this complaint indicates that no patient injury was sustained.
|
| |
|
Additional Manufacturer Narrative
|
|
Balt usa reference: # (b)(4).Investigation pending return of suspected device.The submission of this report or related information to the fda, and its release by fda, does not reflect a conclusion by the party submitting this report of the fda that the report or related information is an admission that the manufacturer, their employees, or the device caused or contributed to the reportable event.
|
| |
|
Search Alerts/Recalls
|