| Model Number |
22-101 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Pericardial Effusion (3271)
|
| Date of Event |
06/24/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
One (b)(6) 2026, a patient underwent a thrombectomy procedure using the flowtriever system of devices.The physician believes the patient developed pericardial effusion due to the size of the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer reference number: (b)(4).The suspect device was not returned to the manufacturer and could not be evaluated.The device identifiers were provided, and the lot history records were reviewed.There were no discrepancies or unusual findings.Additionally, a complaint history review was conducted, and no similar issues have been reported against this lot number.The stryker peripheral vascular medial affairs team reviewed the available information and could not rule out that the peripheral vascular device may have contributed to the reported event.Device labelling lists pericardial effusion as a potential adverse event associated with use.
|
| |
|