• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Allergan (Costa Rica) Inspira Smooth Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Allergan (Costa Rica) Inspira Smooth Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled Back to Search Results
Catalog Number SSF-335
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported the implant infection was positive for the bacteria "pseudomonas aeruginosa".After irrigation, it was reported ¿no infection present¿.This record is for the right side.The device was explanted.
 
Event or Problem Description
Healthcare professional reported an ¿implant infection¿.This record is for the right side.The device was explanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: h6.Further investigation results: with the information collected during the investigation, there is enough evidence to support that units involved in this work order were manufactured under controlled conditions in accordance with abbvie¿s procedures and were no defects/problems/issues found in the documents related to the reported event.Seal strength was successfully completed and results were acceptable.Environmental monitoring results were found to be acceptable, and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.One additional complaint was noted for devices sterilized under sterilization run 30051096.However, it is unlikely that the events related to the reported infection are associated with the sterilization methods utilized.The terminal sterilization cycle used by abbvie was developed and validated to achieve a high level of sterility assurance.The sterilization cycle has been validated to ensure that the chance of a non-sterile device is less than one in a million.Considering that there is no adverse trend for this type of event, no additional actions are deemed required at this time.The issue with infection will continue to be monitored and corrective actions will be taken in the future if deemed appropriate.Given the fact that the cause of the infection cannot be specifically associated with the manufacturing process, and there is not an adverse trend for this type of event, and no er/ ncr(s) were identified during the investigation associated with this lot and the complaint, no corrective action is deemed necessary at this time.
 
Additional Manufacturer Narrative
This report was impacted by emdr scheduled maintenance occurring july 22-27, 2026.The event of infection is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inspira Smooth Silicone Gel Filled Breast Implant
Common Device Name
prosthesis, breast, noninflatable, internal, silicone gel-filled
Manufacturer (Section D)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
+1(847)936-6324
MDR Report Key50015624
Report Number9617229-2026-14053
Device Sequence Number15836434
Product Code FTR
UDI-Device Identifier10888628033122
UDI-Public(01)10888628033122(17)Unknown
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSSF-335
Device Lot Number1212511
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 06/29/2026
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date07/27/2026
Supplement Report FDA Received Date08/19/2026
09/03/2026
Date Device Manufactured10/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight136 LB
-
-