| Model Number |
6143.1001 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the surgery was an l2-s1 and we had two screws misplaced while using gps with the end effector.The first screw was revised using freehand navigation.The second one was left out after trying to replace with freehand navigation but was unsuccessful.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation revealed that there was no system malfunction.It was reported that a system malfunction was observed during the case.Review of the case logs revealed that there was skiving, which resulted in excessive deflection at both the misplaced implants.The case was completed with no adverse effects to the patient reported.The severity observed did not exceed the anticipated severity.The observed overall risk level is low, which matches the observed anticipated risk level; therefore, the overall risk of the system has been maintained and there is no further investigation required.The cause of the reported issue can be traced to user technique.
|
| |
|
Search Alerts/Recalls
|