| Model Number |
CARDINAL GLENNON MO 6 |
| Medical Device Problem Code |
Component Misassembled (4004)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
See initial report.
|
| |
|
Additional Manufacturer Narrative
|
|
H11:
photos were provided which confirmed the reported condition.Complaints database review revealed other 3 similar events occurred at the same customer facility, impacting the same perfusion tubing system product code.No concerning trend was highlighted for reported failure mode.Based on investigation findings, the event was traced back to an operator error in preforming the manual assembly.Manufacturing personnel re-training is required.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
| |
|