| Model Number |
CA561TS |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A remstar auto device was returned in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no report of serious patient harm or injury.There was no report of medical intervention.During the evaluation of the device at the third-party service center, the device was visually inspected and found a burned connector.The device was forwarded to the riql for further investigation.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Event or Problem Description
|
|
A remstar auto device was returned in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no report of serious patient harm or injury.There was no report of medical intervention.During the evaluation of the device at the third-party service center, the device was visually inspected and found a burned connector.The device was forwarded to the pil for further investigation.During the investigation, the pil observed top enclosure and user interface knob missing.The pil also observed dust-like contamination on the air outlet port, air inlet, printed circuit assembly, on top of the blower box lid and surrounding area, on top of the blower motor and inside of the blower box, bottom enclosure, and the air inlet seal.Thermal event on humidifier harness connector and pca at j9 was observed.The sound abatement foam was intact and had dust-like contamination on top of it.The source of contamination was most likely external to the device.Pil was able to confirm the complaint of burnt humidifier base cable.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of thermal issue is accommodated in the product risk management file as being linked to hazard with description fire, flame, smoke.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no further indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.
|
| |
|
Search Alerts/Recalls
|
|
|