Abbott Medical Navitor Transcatheter Aortic Valve; Aortic Valve, Prosthesis, Percutaneously Delivered Back to Search Results
Model Number NVTR-29
Medical Device Problem Codes Material Split, Cut or Torn (4008); Intravalvular regurgitation (4068)
Health Effect - Clinical Code Aortic Valve Insufficiency/ Regurgitation (4450)
Date of Event 05/13/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
(b)(4) vista nova, patient site id: (b)(6).\ \ it was reported that on (b)(6) 2024, a 29mm navitor valve was chosen for implant.It was then reported on (b)(6) 2026, patient presented with severe aortic regurgitation.A transesophageal echocardiogram on (b)(6) 2026 confirmed presence of central aortic regurgitation due to a leaflet tear.No intervention or medical treatment was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Navitor Transcatheter Aortic Valve
Common Device Name
Aortic Valve, Prosthesis, Percutaneously Delivered
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
St. Jude Medical #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50019961
Report Number2135147-2026-05090
Device Sequence Number9662364
Product Code NPT
UDI-Device Identifier05415067031600
UDI-Public(01)05415067031600(17)260418
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P190023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/18/2026
Device Model NumberNVTR-29
Device Catalogue NumberNVTR-29
Device Lot Number9234164
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2026
Initial Report FDA Received Date07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
Patient Weight125 LB
-
-