| Catalog Number |
MX61260 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hematoma (1884)
|
| Date of Event |
06/29/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported in the move study, a patient experienced groin swelling after the use of three 6-12f mynx control venous vascular closure device (vcd) in the right groin.An ultrasound was completed on follow up visit showing a 3.1 x 1.3 x 0.7 complex debris filled collection which is likely hematoma.Medication was provided as treatment.The device was used after an atrial fibrillation (afib) ablation using a non-cordis device.Venous access was gained with the use of ultrasound.One 7f, one 8f and one 10f sheath were use with approximately 6mm in between sites.Hemostasis was maintained after 5 minutes.The devices were successfully deployed.The patient received 23000 units of intraprocedural heparin and 50mg of protamine was administered.The activated clotting time (act) was 383 seconds.Procedural time was 33 minutes, and time to ambulation was 2.28 hours.The device was stored and prepped according to the ifu.The deployer was trained in the use of the device.The devices will not be returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to system limitations, section h6 required to input type of investigation, investigation findings, and investigation conclusions for an initial report.Pending additional information to complete investigation.Additional information is pending and will be submitted within 30 days upon receipt.
|
| |
|