• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Bard Peripheral Vascular, Inc. Maxcore; Biopsy instrument

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Bard Peripheral Vascular, Inc. Maxcore; Biopsy instrument Back to Search Results
Catalog Number MC1610
Medical Device Problem Codes Difficult to Remove (1528); Failure to Fire (2610); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2026, a patient underwent an ultrasound guided breast biopsy procedure through hard tissue density using the maxcore instrument.During the procedure, the device needle was found to be bent.It was further the firing button got stuck a bit and the needle was difficult to remove from the patient.No coaxial needle was used.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: as the lot number for the device was provided, a review of the device history records was performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Maxcore
Common Device Name
Biopsy instrument
Manufacturer (Section D)
Bard Peripheral Vascular, Inc.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
Bard Reynosa S.A. de C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key50021351
Report Number2020394-2026-01906
Device Sequence Number21464805
Product Code KNW
UDI-Device Identifier00801741084447
UDI-Public(01)00801741084447(11)250416(17)280131(10)REKP1221
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K133948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMC1610
Device Lot NumberREKP1221
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2026
Initial Report FDA Received Date07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-