Catalog Number MX61260
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 06/29/2026
Type of Reportable Event Serious Injury
Event or Problem Description
As reported in the move study, a patient experienced groin swelling after the use of three 6-12f mynx control venous vascular closure device (vcd) in the right groin.An ultrasound was completed on follow up visit showing a 3.1 x 1.3 x 0.7 complex debris filled collection which is likely hematoma.Medication was provided as treatment.The device was used after an atrial fibrillation (afib) ablation using a non-cordis device.Venous access was gained with the use of ultrasound.One 7f, one 8f and one 10f sheath were use with approximately 6mm in between sites.Hemostasis was maintained after 5 minutes.The devices were successfully deployed.The patient received 23000 units of intraprocedural heparin and 50mg of protamine was administered.The activated clotting time (act) was 383 seconds.Procedural time was 33 minutes, and time to ambulation was 2.28 hours.The device was stored and prepped according to the ifu.The deployer was trained in the use of the device.The devices will not be returned for evaluation.
 
Additional Manufacturer Narrative
Due to system limitations, section h6 required to input type of investigation, investigation findings, and investigation conclusions for an initial report.Pending additional information to complete investigation.Additional information is pending and will be submitted within 30 days upon receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYNX CONTROL VENOUS
Common Device Name
Device, Hemostasis, Vascular
Manufacturer (Section D)
Cordis US Corp
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
Cordis US Corp.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
5452 betsy ross drive
santa clara, CA 95054
+1(786)313-8372
MDR Report Key50021510
Report Number3004939290-2026-00985
Device Sequence Number9662413
Product Code MGB
UDI-Device Identifier10705032090278
UDI-Public(01)10705032090278(17)280306(10)F2606504
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P040044
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberMX61260
Device Lot NumberF2606504
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2026
Supplement Date Received by Manufacturer08/04/2026
Initial Report FDA Received Date07/27/2026
Supplement Report FDA Received Date08/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/06/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
unknown sheath 7F/10F/9F
Patient Age70 YR
Patient SexMale
Patient Weight208 LB
-
-