Catalog Number ERSD-180Q
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As stated in the literature, ¿a number of risk factors for rupture have been identified; the most common cause is surgical instrument damage.¿ (handel et al., 2013).Additionally, rupture is a common and known reason for complaints, and it is clearly characterized in the product dfu included with the implant, as stated above.Additionally, no cases of rupture because of product failure have been reported to establishment labs ever.Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health the instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: ¿breast implants can potentially remain intact for decades in the body, but all such devices will fail at some point.Breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation, but it is most likely to occur; the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress, and weakening during implantation, age and design of the implant, submuscular rather than sub-glandular location, the occurrence of post-operatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture¿.Establishment labs requests the unit to confirm the event and to perform the corresponding testing (including but not limited to: revision and characterization under the microscope of the rupture section, mechanical testing according to approved standards, etc.).For this specific case, no response was received regarding device return.Breast implants are not considered lifetime devices, and implant rupture is a risk associated with breast surgery, which can occur at any time.
 
Event or Problem Description
Mexico.Allegedly, a patient attended a gynecological follow-up appointment during which an ultrasound and mammogram were ordered; imaging findings suggested a possible implant rupture, and the report was attached to the medical record.The investigation is ongoing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTIVA IMPLANTS
Common Device Name
Ergonomix Round SilkSurface with Qid
Manufacturer (Section D)
Establishment Labs S.A.
building b-25, coyol free zone
alajuela,alajuela 20113
CS  20113
Manufacturer (Section G)
Establishment Labs S.A.
building b-25, coyol free zone
alajuela,alajuela 20113
CS   20113
Manufacturer Contact
susana hidalgo
building b-25, coyol free zone
alajuela,alajuela 20113
CS   20113
+1(800)355-4164
MDR Report Key50021690
Report Number3012883202-2026-02672
Device Sequence Number13406460
Product Code FTR
UDI-Device Identifier07445161012081
UDI-Public(01)07445161012081(11)220106(17)270106
Combination Product (Y/N)N
Initial Reporter CountryMX
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberERSD-180Q
Device Lot Number21121499
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2026
Initial Report FDA Received Date07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-