| Catalog Number |
0117080 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Adhesion(s) (1695)
|
| Date of Event |
03/21/2011
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Per information provided:
(b)(6) 2009 - patient was diagnosed with left inguinal hernia thereby underwent open repair with implant of kugel patch (device 1).Per op notes, ¿a transverse incision was made above the midportion of the left inguinal ligament.An indirect hernia sac was identified and reduced back into the peritoneum.A kugel patch (device 1) was placed into the preperitoneal pocket and secured with sutures.¿
(b)(6) 2010 - patient was diagnosed with recurrent left inguinal hernia, indirect type thereby underwent open repair with implant of perfix light plug (device 2).Per op notes, ¿an inguinal incision was made.An indirect sac was identified and cleared from medial to lateral.A perfix light plug (device 2) was placed into the preperitoneal space.The previously placed mesh (device 1) could be palpated along the cephalad border of the defect and the perfix light plug (device 2) was secured to this piece of mesh (device 1).¿
(b)(6) 2011 - patient was diagnosed with prostate cancer thereby underwent radical prostatectomy.Per op notes, ¿a midline incision was made.Left pelvic space was created and noted to have a mesh (device 1 and 2) which extends from the abdominal wall and the peritoneum were adherent to this mesh (device 1 and 2).The mesh (device 1 and 2) was overlying the left iliac vessels and the plane leriche was not able to be developed.Some thickening related to the mesh (device 1 and 2) in the left pelvic space was noted.A radical prostatectomy was performed.¿.
|
| |
|
Additional Manufacturer Narrative
|
|
Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.This mdr represents the perfix light plug (device 2).An additional supplemental emdr was submitted to represent the kugel patch (device 1).Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|