• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. Bard Inc. (Covington) -1018233 Bardex® Lubricath® Foley Catheter; Latex Foley Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. Bard Inc. (Covington) -1018233 Bardex® Lubricath® Foley Catheter; Latex Foley Catheter Back to Search Results
Catalog Number 0165L20
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer that the foley catheter balloon leaked.The catheter was placed; however, it came out after 5 minutes.Upon inspection, the balloon was inflated and found to be leaking, resulting in deflation.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.A potential root cause for this failure mode could be due to thin rubberized wall causing collapsed or pinched inflation lumen under the balloon or along shaft.Labeling review not required.The dhr review could not be performed without a lot number.Therefore, no additional action is required at this time.Correction: d,e udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Bardex® Lubricath® Foley Catheter
Common Device Name
Latex Foley Catheter
Manufacturer (Section D)
C.R. Bard Inc. (Covington) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. Bard Inc. (Covington) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
+1(770)784-6100
MDR Report Key50022033
Report Number1018233-2026-04866
Device Sequence Number9662421
Product Code EZC
UDI-Device Identifier00801741017506
UDI-Public(01)00801741017506
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0165L20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2026
Supplement Date Received by Manufacturer08/26/2026
Initial Report FDA Received Date07/27/2026
Supplement Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-