| Catalog Number |
0165L20 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that the foley catheter balloon leaked.The catheter was placed; however, it came out after 5 minutes.Upon inspection, the balloon was inflated and found to be leaking, resulting in deflation.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.A potential root cause for this failure mode could be due to thin rubberized wall causing collapsed or pinched inflation lumen under the balloon or along shaft.Labeling review not required.The dhr review could not be performed without a lot number.Therefore, no additional action is required at this time.Correction: d,e
udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|