MAUDE Adverse Event Report: Unomedical Devices S.A de C.V Comfort; Set, Administration, Intravascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Unomedical Devices S.A de C.V Comfort; Set, Administration, Intravascular Back to Search Results
Model Number MMT-377A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 06/22/2026
Type of Reportable Event Serious Injury
Event or Problem Description
To date, additional patient or event details were received which have been added in h11 (addl mfg narrative).
 
Event or Problem Description
It was reported that the patient experienced hyperglycemia of 316 mg/dl which treated with manual bolus via insulin pump, manual injection and insulin pen on (b)(6) 2026.
 
Additional Manufacturer Narrative
Imdrf investigation findings code: code c1902 is not available in database to capture device problem excluded.Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary.Complaint investigation results.A complaint investigation was conducted based on the event description and the assigned malfunction code no malfunction based on complaint information.Based on the available information, the event is consistent with misuse, user-related factors, or no malfunction alleged.In accordance with appendix of wi complaint handling, the need for additional investigation activities, including product testing, or electronic quality management system (eqms) search, was evaluated and determined not to be warranted for this record.However, as a lot number was provided and the event involved a serious injury, a batch record / medical device file review was conducted for lot 6016664 to assess any potential manufacturing-related issues.Batch record / medical device file review: the batch record / medical device file was reviewed.All required in-process and final tests were completed and met specified requirements.No deviations were identified, and no related maintenance activities were recorded.Conclusion: batch record / medical device file review supports compliance with manufacturing and quality requirements; no issues noted.Capa determination assessment, conclusion summary of complaint investigation: based on the investigation, no further investigation is required.The record will be closed and monitored through tracking and trending per work instruction (wi) (monthly trips and alerts).Complaint unconfirmed: following investigation, the reported failure could not be confirmed for this complaint.Fda registration number.Reporting site: 8021545.Manufacturing site: 3003442380.
 
Additional Manufacturer Narrative
Based on the available information, this event is deemed to be a serious injury.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number.Reporting site: 8021545.Manufacturing site: 3003442380.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Comfort
Common Device Name
Set, Administration, Intravascular
Manufacturer (Section D)
Unomedical Devices S.A de C.V
ave.fomento industrial l9 m3
parque industrial del norte
reynosa,tamps. 88736
MX  88736
Manufacturer (Section G)
Unomedical Devices S.A de C.V
ave.fomento industrial l9 m3
parque industrial del norte
reynosa,tamps. 88736
MX   88736
Manufacturer Contact
pamela meadows
aaholmvej 1 - 3
osted
lejre 4320
DA   4320
+1(603)247-0255
MDR Report Key50022995
Report Number3003442380-2026-57483
Device Sequence Number18543527
Product Code FPA
UDI-Device Identifier05705244019522
UDI-Public05705244019522
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K162812
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-377A
Device Lot Number6016664
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2026
Supplement Date Received by Manufacturer08/04/2026
Initial Report FDA Received Date07/27/2026
Supplement Report FDA Received Date08/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-