Institut Straumann AG CIM ?3.3mm ND, ZLA 10mm, AH5.5mm, ZrO2; Endosseous Dental Implant Back to Search Results
Catalog Number 031.012S
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Fibrosis (3167)
Date of Event 06/26/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2024 in fdi 24.On (b)(6) 2026, loss of osseointegration was verified.The device was forwarded to the manufacturer.At the event the patient experienced: mobility.No further patient complications were reported.
 
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.? the loss of an endosseous dental implant after successful osseointegration and restoration is a known inherent risk of the procedure.Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: 1.Absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia.2.Absence of a recurrent peri-implant infection with suppuration.3.Absence of implant mobility.4.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CIM ?3.3mm ND, ZLA 10mm, AH5.5mm, ZrO2
Common Device Name
Endosseous Dental Implant
Manufacturer (Section D)
Institut Straumann AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
Institut Straumann AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
+1(978)747-2509
MDR Report Key50023539
Report Number0009613348-2026-011337
Device Sequence Number9616116
Product Code DZE
UDI-Device Identifier07630031713834
UDI-Public07630031713834
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K151328
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 07/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/02/2025
Device Catalogue Number031.012S
Device Lot NumberAZA52
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/30/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2026
Initial Report FDA Received Date07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/02/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age84 YR
Patient SexMale
-
-