It was reported that on (b)(6) 2026, a 27mm navitor vision valve was selected for implant using a flexnav delivery system (ds).Pre-implant balloon aortic valvuloplasty (pre-bav) was performed by using a 23mm, non-abbott balloon.During the procedure, the valve was recaptured as it was positioned higher on the left-coronary cusp (lcc).On the second attempt, the patient became hypotensive and the valve was quickly deployed in a deeper position.On echocardiogram (echo) showed pericardial effusion; the physician(s) conclude the presence of cardiac tamponade and obstructive shock.Pericardiocentesis was performed to resolve the event and there was no clinically significant delay reported.The patient became stable and the user suspects that the adverse event might be due to the manipulation of non-abbot wire.The patient had an extended hospitalization and in a stable condition, scheduled (waiting) for discharge.
|