DePuy Ireland - 9616671 ATTUNE; Tibial Trials Back to Search Results
Catalog Number 254505262
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the trials have green discoloration and need to be replaced.The event did not happen in surgery.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: h3, h6: investigation summary it was reported that the trials have green discoloration and need to be replaced.The event did not happen in surgery.The product was returned to depuy synthes for evaluation.Visual inspection of the returned device found what appears to be a discoloration of the device or a green foreign substance attached/stained into the device surface.No other cosmetic anomaly was identified on the evidence provided.Although there is no certainty of the origin of the foreign substance/matter, potential cause can be traced to improper cleaning/maintenance of the device.As per ifu, the following precautions need to be followed during the cleaning/maintenance process: 1) prepare an enzymatic neutral ph cleaning solution in accordance with the manufacturer¿s instructions; 2) soak soiled instruments per manufacturer¿s instructions (minimum of 5 minutes) in the enzymatic solution; 3) use a soft bristle brush to remove all traces of blood and debris; pay close attention to any hard-to-reach areas, textured surfaces, or crevices; 4) rinse the instrument thoroughly with warm tap water; and 5) if possible, ultrasonically clean the instrument for 10 minutes in neutral ph detergent, prepared in accordance with the manufacturer¿s instructions and rinse the instrument thoroughly with warm tap water; and 6) dry the instrument immediately after final rinse.Properly handling and¿attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune fb ps trl would have contributed to the complained device issue.¿ based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE
Common Device Name
Tibial Trials
Manufacturer (Section D)
DePuy Ireland - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
+1(908)386-3534
MDR Report Key50025157
Report Number1818910-2026-12669
Device Sequence Number10579032
Product Code HWT
UDI-Device Identifier10603295478881
UDI-Public(01)10603295478881(10)BFA1BL9
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254505262
Device Lot NumberBFA1BL9
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/30/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 07/24/2026
Initial Report FDA Received Date07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-