It was reported that a monofocal toric intraocular lens (iol) was explanted from a patient's left eye since the patient experienced a significant change in corneal astigmatism from preoperative status, resulting in diplopia.The patient's best spectacle-corrected visual acuity improved from 20/60 preoperatively to 20/20 postoperatively.The issue did not affect patient's daily activities.No vitrectomy or incision enlargement required, the patient did not seek medical attention and no medication outside of standard of care was prescribed.The replacement lens implanted was the same model and diopter but different cylinder (diu150, 19.5d).The patient is temporarily impaired.Direction for use were followed.The lens involved is not available for return.No additional information was provided.
|
|
Section a4: weight: unknown, information was asked but not provided.Device evaluation:
product evaluation was not performed because the product has not been received.The complaint issue reported could not be confirmed.Manufacturing record review:
the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search of complaints related to this production order (po) was performed.The search revealed one additional complaint folder was received for this po.However, complaint issues reported are not related.Therefore, no further actions are required.Conclusion:
based on the investigation, no malfunction or product deficiency was identified.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.This report was impacted by emdr scheduled maintenance occurring july 22-27, 2026.
|