the mayfield infinity skull clamp (a1114) was returned for evaluation:
device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the inspection of the skull clamp shows the skull clamp has rotational movement in its swivel lock assembly and a residue buildup is present.Which does not guarantee proper fixation of the patient's skull.For the adjustment of the lock assembly, new components need to be added to replace worn internal parts.To resolve the issues, the skull clamps internal parts were replaced to adjust the unit according to the manufacturer specifications and to clean the skull clamp's swivel lock assembly.After completing the reassembly, including the adjustment, the skull clamp was subjected to a successful function test and was confirmed to meet all manufacturer specifications.Root cause - the complaint is confirmed via inspection of the unit.The clamp had rotational movement in its swivel lock assembly and residue buildup, requiring cleaning and replacement of worn parts.The probable root cause is routine use and wear.Additionally, improper or suboptimal placement of the skull clamp can continue to movement or slippage of the patient.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|