Integra LifeSciences Corporation OH/USA MAYFIELD INFINITY SKULL CLAMP Back to Search Results
Catalog Number A1114
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11 the mayfield infinity skull clamp (a1114) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the inspection of the skull clamp shows the skull clamp has rotational movement in its swivel lock assembly and a residue buildup is present.Which does not guarantee proper fixation of the patient's skull.For the adjustment of the lock assembly, new components need to be added to replace worn internal parts.To resolve the issues, the skull clamps internal parts were replaced to adjust the unit according to the manufacturer specifications and to clean the skull clamp's swivel lock assembly.After completing the reassembly, including the adjustment, the skull clamp was subjected to a successful function test and was confirmed to meet all manufacturer specifications.Root cause - the complaint is confirmed via inspection of the unit.The clamp had rotational movement in its swivel lock assembly and residue buildup, requiring cleaning and replacement of worn parts.The probable root cause is routine use and wear.Additionally, improper or suboptimal placement of the skull clamp can continue to movement or slippage of the patient.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY SKULL CLAMP
Common Device Name
Mayfield
Manufacturer (Section D)
Integra LifeSciences Corporation OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
Manufacturer (Section G)
Integra LifeSciences Corporation OH/USA
4900 charlemar drive
cincinnati OH 45227
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
+1(609)936-2319
MDR Report Key50025999
Report Number3004608878-2026-00082
Device Sequence Number18543551
Product Code HBL
UDI-Device Identifier10381780253549
UDI-Public10381780253549
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K051440
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Type of Report Initial,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberA1114
Device Lot Number121745/151
Date Returned to Manufacturer07/03/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2026
Supplement Date Received by Manufacturer08/27/2026
Initial Report FDA Received Date07/27/2026
Supplement Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-