MAUDE Adverse Event Report: ZimVie US Corp LLC OSSEOTITE® CERTAIN® 2 IMPLANT 5 X 11.5MM; Dental Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZimVie US Corp LLC OSSEOTITE® CERTAIN® 2 IMPLANT 5 X 11.5MM; Dental Implant Back to Search Results
Catalog Number XIFOSS511
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 06/30/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant at tooth #30 was removed due to bone loss.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: patient weight unknown / not provided a summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® CERTAIN® 2 IMPLANT 5 X 11.5MM
Common Device Name
Dental Implant
Manufacturer (Section D)
ZimVie US Corp LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZimVie US Corp LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
+1(561)971-3230
MDR Report Key50033246
Report Number0001038806-2026-03916
Device Sequence Number15915889
Product Code DZE
UDI-Device Identifier00844868023227
UDI-Public(01)00844868023227(17)270523(10)2022050234
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K100724
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 07/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberXIFOSS511
Device Lot Number2022050234
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/16/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 07/07/2026
Initial Report FDA Received Date07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/23/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
-
-