Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: deflation.This report was impacted by emdr scheduled maintenance occurring july 22-27, 2026.
|
|
Device evaluation: the device related to the reported events deflation was received on september 03, 2026 with lot number 1270784.Per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- deflation: not observed through leak test inspection.None of the other observations performed during the device analysis creases and partial delamination of diaphragm valve are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Additional, changed, and/or corrected data: d.9., h.2., h.3., h.6.
|