Visual inspection
during the investigation, scratches on the outer housing of the progav 2.0 were determined.Permeability test
a permeability test has shown that the valve is permeable.Computer controlled test
to investigate the claim of over- drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in the horizontal position.The results show that the valve operates within the accepted tolerances in the horizontal position.Adjustment test
the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test
the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection
after dismantling of the valve, no organic deposits were found.Result
based on the test results, no functional deviations or other abnormalities were found in the progav 2.0.The valve is operating within its specifications.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|