| Catalog Number |
035.8206S |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/07/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The clinician reports that the day the implant was placed in ada 30, failure occurred upon insertion.Details of surgery: primary stability not achieved and immediate extraction site.The device was forwarded to the manufacturer.There were no reported patient injuries or complications.
|
| |
|
Additional Manufacturer Narrative
|
|
The batch number could not be verified due to incomplete or missing
information and / or product from the customer.Our manufacturing
q-system assures that production and process controls are in place to
ensure that batches confirm to the applicable specifications before they
are distributed.The removal of a dental implant during surgery without the replacement
of another dental implant is a known inherent risk of the procedure due
to either lack of primary stability of the implant (patienor procedure
related).It may also include the removal of an implant after
osseointegration due to either the clinician's or patient¿s decision.The manufacturer¿s trend analysis confirms that usually procedural
errors and/or patient's condition contribute to the event.
|
| |
|