MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VIGILANT? EL ICD VR; Implantable Cardioverter Defibrillator (ICD)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VIGILANT? EL ICD VR; Implantable Cardioverter Defibrillator (ICD) Back to Search Results
Model Number D232
Medical Device Problem Code Signal Artifact/Noise (1036)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/25/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited noisy signals.Technical services (ts) reviewed the stored ventricular tachycardia (vt) episodes and noted noise without oversensing.Ts suspected diaphragmatic noise and recommended to continue to monitor.No adverse patient effects were reported.This icd remains in service.Additional information was received that this icd was successfully explanted and replaced.No additional adverse patient effects were reported outside the revision procedure.At this time, this product has not been returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT? EL ICD VR
Common Device Name
Implantable Cardioverter Defibrillator (ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
+1(651)582-6168
MDR Report Key50040887
Report Number2124215-2026-40098
Device Sequence Number13222628
Product Code LWS
UDI-Device Identifier00802526587788
UDI-Public(01)00802526587788(17)241212
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/12/2024
Device Model NumberD232
Device Catalogue NumberD232
Device Lot Number310361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2026
Initial Report FDA Received Date07/27/2026
Date Device Manufactured12/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age64 YR
Patient SexMale
-
-