| Catalog Number |
68HP-550 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "deflation".This record is for the right side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: deflation.This report was impacted by emdr scheduled maintenance occurring july 22-27, 2026.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation".This recod is for the right side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- deflation: observed an opening through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases, stress marks and wear abrasion are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Search Alerts/Recalls
|