| Catalog Number |
SCM-295 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Foreign Body Reaction (1868)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "exposure".This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report was impacted by emdr scheduled maintenance occurring july 22-27, 2026.A review of the device history record has been completed.No deviations or non-conformances noted.The event of "exposure" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: exposure.
|
| |
|
Search Alerts/Recalls
|