• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Institut Straumann AG SP ? 3.3mm RN, SLA? 14mm, Ti, Loxim; Endosseous Dental Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Institut Straumann AG SP ? 3.3mm RN, SLA? 14mm, Ti, Loxim; Endosseous Dental Implant Back to Search Results
Catalog Number 043.914S
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2022 in fdi 11.On (b)(6) 2026, loss of osseointegration was verified.The device was forwarded to the manufacturer.There were no patient operative or post-operative complications reported.
 
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.? the loss of an endosseous dental implant after successful osseointegration and restoration is a known inherent risk of the procedure.Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: 1.Absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia 2.Absence of a recurrent peri-implant infection with suppuration 3.Absence of implant mobility 4.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SP ? 3.3mm RN, SLA? 14mm, Ti, Loxim
Common Device Name
Endosseous Dental Implant
Manufacturer (Section D)
Institut Straumann AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
Institut Straumann AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
+1(978)747-2509
MDR Report Key50043923
Report Number0009613348-2026-011655
Device Sequence Number5558424
Product Code DZE
UDI-Device Identifier07630031719485
UDI-Public07630031719485
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K033922
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 07/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/13/2026
Device Catalogue Number043.914S
Device Lot NumberENX43
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/01/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2026
Initial Report FDA Received Date07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexMale
-
-