| Model Number |
IO-501 |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
07/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient was having issues with their device.As a result, their oxygen levels had decreased.The patient did not have a backup device at the time of the event.The patient was hospitalized for 3 weeks and had 3 weeks of rehab.The patient has received their replacement device and does not have any ongoing complications due to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event is estimated.The unit was received with a reported oxygen error, confirmed with the contaminated zeolite.Incoming internal 02 measured 89.6% and external 02 measured 89.1 %, both below specification.Furthermore, the psa cycle (according to the data log) being high (9) is an indication that the zeolite is getting contaminated by moisture.Also damaged compressor mounts due to compressor vibration.
|
| |
|