| Catalog Number |
P45-191-5045 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
06/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that half pin broke while inserting during surgery and its debris is still within the t-handle.The other portion of the half pin is implanted in the patient.The break occurred at the very end of the procedure and caused no delay or harm/impact to the patient.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This mdr is submitted late after remediation efforts and a retrospective review of complaint reportability.The reported event was confirmed via product return.Visual examination confirmed device is fractured.Review of the device history records identified no deviations or anomalies during manufacturing.A probably root cause is excessive torque, although this is not confirmed; root cause is not established.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|