• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Inogen, Inc. Inogen One G3 High Flow Portable Oxygen Concentrator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Inogen, Inc. Inogen One G3 High Flow Portable Oxygen Concentrator Back to Search Results
Model Number IO-300
Medical Device Problem Codes Premature Discharge of Battery (1057); Failure to Charge (1085)
Health Effect - Clinical Code Syncope/Fainting (4411)
Date of Event 09/13/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the device was not charging while plugged in at a court house and the device battery was also not lasting as long as expected.As a result, the device ran out of power and the patient passed out due to being without oxygen.The patient was taken to the hospital and put on life support.The patient has since been released and received their replacement device.
 
Additional Manufacturer Narrative
B3: date of event is estimated.There is no confirmation for the return reason of battery error.Zeolite is found contaminated which possibly caused when it absorbs the moisture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inogen One G3 High Flow Portable Oxygen Concentrator
Common Device Name
Inogen One G3 High Flow Portable Oxygen Concentrator
Manufacturer (Section D)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer (Section G)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer Contact
greg porter
600 shiloh rd
plano, TX 75074
MDR Report Key50046428
Report Number3004672275-2026-00217
Device Sequence Number21202631
Product Code CAW
UDI-Device Identifier00817131020063
UDI-Public(01)00817131020063
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K032818
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 07/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberIO-300
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/15/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2026
Initial Report FDA Received Date07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening;
Patient SexUnknown
-
-