• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO Puerto Rico Mfg. Inc. SENSAR IOL; Intraocular Lens

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO Puerto Rico Mfg. Inc. SENSAR IOL; Intraocular Lens Back to Search Results
Model Number AR40e
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/19/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that two non-preloaded monofocal intraocular lenses (iols) were associated with complications during a surgical procedure involving the patient's right eye.The first lens (ar40c +23.5, sn (b)(6)) was the initial lens used.During implantation, a superior haptic was reported to be broken.The lens was not fully inserted into the patient's eye when the issue was identified.The second lens (ar40c +24.0, sn (b)(6)) was then implanted; however, disinsertion of the posterior haptic was observed.This lens was fully inserted into the patient's eye and subsequently removed.Additional information indicated that a capsular rupture occurred and that an unplanned anterior vitrectomy and incision enlargement were performed during the procedure.However, it is unknown at what point during the procedure the capsular rupture occurred or whether the vitrectomy and incision enlargement were associated with the first lens, the second lens, or another aspect of the procedure.No sutures were required.No replacement lens was implanted following removal of the lenses, and the patient remained aphakic.No cartridge tip damage or use error was reported.It was indicated that the device did not cause or contribute to the event.No further information was available.This mdr report captures the issues reported with the first lens.A separate report will be submitted for the issues reported with the second lens.
 
Additional Manufacturer Narrative
Section a4, a5 and a6: unknown, information was requested but not provided.Section d6a.If implanted, give date: unknown, information not provided.Section d6b.If explanted, give date: not applicable, as there is no indication that the device was explanted.Section e1: initial reporter establishment name: unknown, information not provided.Section e1: initial reporter's address, city, state and zip code: unknown, information not provided.Section e1: phone number: (b)(6) device evaluation: the product testing could not be performed as the product was not returned for evaluation.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search of complaints related to this production order (po) was performed.The search revealed that no other complaints were received for this production order number.Conclusion: based on the investigation, no malfunction or product deficiency was identified.Section h6: health effect - impact code: 4625 - additional surgery (incision enlarged & unplanned vitrectomy) section h6: health effect - clinical code: 4581 incision enlarged.Attempts were made to obtain additional information regarding the event; however, the information was not provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.This report was impacted by emdr scheduled maintenance occurring july 22-27, 2026.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR IOL
Common Device Name
Intraocular Lens
Manufacturer (Section D)
AMO Puerto Rico Mfg. Inc.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
+1(714)247-8552
MDR Report Key50047507
Report Number3012236936-2026-000215
Device Sequence Number10628377
Product Code HQL
UDI-Device Identifier05050474502178
UDI-Public(01)05050474502178(17)310113
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAR40e
Device Catalogue NumberAR40E00235
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2026
Initial Report FDA Received Date07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/13/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-