• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR Surgical Company EVO/EVO + VISIAN Implantable Collamer Lens (ICL); Phakic Toric Intraocular Lens

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR Surgical Company EVO/EVO + VISIAN Implantable Collamer Lens (ICL); Phakic Toric Intraocular Lens Back to Search Results
Model Number VTICM5 12.1 (-11.00/ 2.00)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation.Lens was repositioned but that did not resolve the problem.In a separate surgery the lens was exchanged with a longer length and the problem was resolved.The cause of the event was reported as unknown.
 
Additional Manufacturer Narrative
H11- manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.H6- investigation type code (b): 4110- lens work order search-no similar complaint event(s) within associated lots were found.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVO/EVO + VISIAN Implantable Collamer Lens (ICL)
Common Device Name
Phakic Toric Intraocular Lens
Manufacturer (Section D)
STAAR Surgical Company
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR Surgical Company
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
MDR Report Key50047866
Report Number2023826-2026-01375
Device Sequence Number18514220
Product Code QCB
UDI-Device Identifier00840311305049
UDI-Public(01)00840311305049(11)250825(17)280731
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVTICM5 12.1 (-11.00/ 2.00)
Device Catalogue NumberVTICM5 12.1
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/11/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2026
Supplement Date Received by Manufacturer08/28/2026
Initial Report FDA Received Date07/27/2026
Supplement Report FDA Received Date09/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-