| Model Number |
SONATA |
| Medical Device Problem Codes |
Impedance Problem (2950); Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
The explanted device was received for investigation.The user was re-implanted.
|
| |
|
Event or Problem Description
|
|
The explanted device was received for investigation.According to the derf a malfunction of the electrode was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.According to information from the derf, the active electrode migrated partially out of cochlea.The irc states a full insertion of the electrode at the time of implantation, therefore, electrode migration out of cochlea seems to be the cause of explantation.Mechanical damages found during investigation are attributable to the removal surgery.This is a final report.
|
| |
|