Model Number 3600-100
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: a2, b4, g1, g2, g3, g6, h2, h6, h11 the complaint reported that an audible leak was heard during use of an oasis drain.The reporter indicated that the same issue occurred with a second drain from the same lot, which was reported separately.The product associated with this complaint was not returned for evaluation.However, testing of the related returned unit identified no damage, and the product functioned as intended.No audible leak or abnormal sound could be reproduced during evaluation.The complaint did not specify the source of the reported sound.Potential sources external to the device, such as the patient line connection or wall suction equipment, could not be evaluated and may have contributed to the reported observation.A device history record review found no manufacturing non conformances, and review of the instructions for use confirmed that adequate setup and operating instructions are provided.Complaint trending found that the actual occurrence level did not exceed the anticipated occurrence level.Based on the available information, neither the reported audible leak nor a product-related non conformance could be confirmed.No evidence of a manufacturing or product deficiency was identified, and the root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Oasis
Common Device Name
Bottle, Collection, Vacuum
Manufacturer (Section D)
Atrium Medical Corporation
40 continental blvd
merrimack NH
Manufacturer (Section G)
Atrium Medical Corporation
40 continental blvd
merrimack NH
Manufacturer Contact
lori gosselin
40 continental blvd
merrimack, NH 
MDR Report Key50054369
Report Number3011175548-2026-000008
Device Sequence Number18521134
Product Code KDQ
UDI-Device Identifier00650862110012
UDI-Public(01)00650862110012(10)522738(17)280810
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K043140
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3600-100
Device Catalogue Number3600-100
Device Lot Number522738
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2026
Supplement Date Received by Manufacturer09/02/2026
Initial Report FDA Received Date07/28/2026
Supplement Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/13/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
Patient Weight239 LB
-
-