Model Number PFTM6.5P-NC
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a nasogastric feeding tube clogged.A new tube was placed.
 
Additional Manufacturer Narrative
This event was initially deemed not reportable based on our reporting strategy active at the time of the alert date; however, following a recent retrospective review of complaints this event was made reportable out of an abundance of caution.The device history record for lot ty241028 was reviewed and the product was produced according to product specifications.The actual complaint product was returned for evaluation.The skives appear fully formed.There is no buildup or occlusion in the skives.The opening at the distal tip is also without occlusion.The defect was not confirmed.The device functioned as intended.All information reasonably known as of 28 jul 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENFit PU Feeding Tube Purple
Common Device Name
DH Nasogastric (NG) Feeding Tubes (NeoMed and NeoConnect)
Manufacturer (Section D)
Avanos Medical Inc.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
Ningbo Tianyi Medical Appliance CO., Ltd
no.788,mozhi north rd
tourism resort, dongqian lake
ningbo zhejiang, cn 31512 1
CH   315121
Manufacturer Contact
tamara cardona
5405 windward parkway
alpharetta, GA 30004
+1(470)448-5487
MDR Report Key50054604
Report Number3011270181-2026-00110
Device Sequence Number10639847
Product Code FPD
UDI-Device Identifier00350770001791
UDI-Public(01)00350770001791(240)120006592(11)260101(17)281231(10)TY241028
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K082238
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPFTM6.5P-NC
Device Catalogue NumberPFTM6.5P-NC
Device Lot NumberTY241028
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/21/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2025
Initial Report FDA Received Date07/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/01/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-